Causes:
1. Pump malfunction
2. Inlet solenoid valve malfunction
3. RO membrane blockage or damage
4. RO membrane housing check valve damage
5. System flushing solenoid valve damage
6. Pretreatment filter blockage
7. Water path blockage before pump
8. Low water temperature
Solutions:
1. Repair or replace the pump.
2. Replace the inlet solenoid valve.
3. Acid wash or replace the RO membrane.
4. Replace the membrane housing check valve.
5. Replace the system flushing solenoid valve.
6. Replace the pretreatment filter.
7. Check the water path before the pump for kinks or blockages.
8. A decrease in inlet water temperature will correspondingly reduce the water production capacity of the RO membrane.
When this machine malfunctions, we recommend that you first try the above methods for diagnosis and troubleshooting. If the problem cannot be resolved, please call our company's service hotline 400-884-6567 immediately, and a UPU ultrapure water engineer will assist you.