Sichuan ULUPURE Ultrapure Technology Co., Ltd.

Study on the Determination of TDS (Total Dissolved Solids) in Water by Electrode Method

Table of Content [Hide]

    I. Introduction


    Total dissolved solids (TDS) are the sum of non-volatile substances obtained after water is filtered and dried at a certain temperature. High concentrations of TDS can give water an unpleasant taste and damage water distribution pipes and equipment, making it an important indicator of water mineralization.


    Currently, two methods are used for detecting TDS in water: drying at 105℃ and drying at 180℃. These are national standard methods, but they are cumbersome, susceptible to interference, difficult to achieve constant weight, and time-consuming, requiring more than a day for a single test.
    A new method, the electrode method, has been developed for determining TDS in water. This method is simple to operate, has a short testing time (a water sample can be tested in just a few minutes), and has yielded satisfactory results with excellent precision and accuracy.


    Tests of the same water sample using the electrode method, the 105℃ drying method, and the 180℃ drying method showed that the difference between the electrode method and the 105℃ drying method was <10%, and the difference between the electrode method and the 180℃ drying method was <5%.


    II. Experimental Section


    1. Instruments and Reagents
    Sension™ 156 Portable Multi-Parameter Analyzer (Hach Instruments, USA)
    Constant Temperature Electric Hot Water Bath (Shanghai Yuejin Medical Instrument Factory)
    101-2-BS Electric Heating Drying Oven (Shanghai Yuejin Medical Instrument Factory)
    Millipure Water System
    10g/L Sodium Carbonate Solution
    Total Soluble Solids Standard (National Standardization Institute)


    2. Method Comparison
    When determining the water sample using the electrode method, the instrument was calibrated with a prepared 200mg/L standard before the water sample was measured. The water sample did not need to be filtered and was measured directly. There were no specific requirements for the sampling volume, as long as the electrode probe could be fully inserted into the water. When determining the water sample using the drying method, the water sample needed to be filtered through a 0.45um pore size filter membrane before measurement, and the sampling volume was accurately 100ml. The same water sample was measured using both the electrode method and the drying method. Result:

    (Unit: mg/L)


    Sample number

    Electrode method measurement value

    Measured value at 105℃

    180℃ measurement value

    The difference from the result measured at 105℃ is %.

    The difference from the result measured at 180℃ is %.

    1

    272

    284

    273

    4.2

    0.4

    2

    273

    281

    266

    2.8

    2.3

    3

    139.9

    152

    138

    8.0

    1.4

    4

    146.2

    158

    144

    7.5

    1.5

    5

    142.2

    149

    146

    4.6

    2.6

    6

    142.5

    152

    148

    6.4

    3.7


    The table shows that: when measuring water samples using the electrode method, the measured values are closer to those measured at 180℃ (drying temperature), with a difference of <5%; when measuring water samples using the electrode method, the measured values are more different from those measured at 105℃ (drying temperature), with some results showing a difference >5%, but <10%. There is no significant difference between the electrode method and the drying method results.


    3. Standard Measurement

    Preparation standard value

    Measured values

    RSD%

    relative error %

    400

    402 402 402 402 402

    0

    0.5

    100

    102.4 102.4 102.3 102.4 102.3 102.4

    0.1

    2.4


    Using a standard calibration instrument with a concentration of 200 mg/L, the prepared standards of 400 mg/L and 100 mg/L were measured. Results: (unit: mg/L)

    As can be seen from the table, the results of the electrode method for determining TDS (total dissolved solids) are accurate and the precision is very good.


    4. Spike Recovery Determination

    A certain amount of standard was added to the water sample using a 200 mg/L standard calibration analyzer, and then the test was performed. The results are as follows:

    (Unit: mg/L)


    water sample baseline

    scalar

    Measured values

    RSD%

    Recovery rate %

    273

    100

    375 375 375 375 375 375

    0

    102

    139.9

    100

    242 242 242 242 242

    0

    102.1

    142.5

    100

    243 243 243 243 243 243

    0

    100.5


    The table shows that the recovery rate for TDS (total dissolved solids) determination using the electrode method ranges from 100.5% to 102.1%.


    III. Conclusion

    The electrode method for determining TDS (total dissolved solids) in water shows no significant difference in results compared to the drying method, indicating that the results are accurate and reliable. The electrode method for determining TDS in water is simple to operate, has a short detection cycle, and can be widely used.

    References
    ULUPURE
    We use cookies to offer you a better browsing experience, analyze site traffic and personalize content. Part of the tracking is necessary to ensure SEO effectiveness,
    By using this site, you agree to our use of cookies. Visit our cookie policy to learn more.
    Reject Accept